Concave cell design for FTIR measurements

Xiang Qin Lin Zhi Xiang Zhang Wan Qun Hu

引用本文: Xiang Qin Lin,  Zhi Xiang Zhang,  Wan Qun Hu. Concave cell design for FTIR measurements[J]. Chinese Chemical Letters, 2011, 22(11): 1339-1342. doi: 10.1016/j.cclet.2011.05.038 shu
Citation:  Xiang Qin Lin,  Zhi Xiang Zhang,  Wan Qun Hu. Concave cell design for FTIR measurements[J]. Chinese Chemical Letters, 2011, 22(11): 1339-1342. doi: 10.1016/j.cclet.2011.05.038 shu

Concave cell design for FTIR measurements

摘要: Wedged and V-shaped cells have advantages on depressing strong absorption bands and keeping the photo windows wide open without absorption block. Alternatively, this work presents a concave cell design, which is constructed by a plano-convex CaF2 lens in combination with a flat plate. Mathematical equations have been provided for data treatment with Matlab programs. The result indicates that the concave cell has advantages similar to the V-shaped cell. Reflection-absorption measurements of ethanol have been conducted on the cell in mid-IR region for demonstration. The spectrum obtained from the concave cell can easily be interpreted to regenerate its normal spectrum, which is fitting well with the spectrum taken in a conventional thin-layer cell. The concave cell is easier to construct and more convenient to operate, willing to have popular usages.

English

  • 加载中
计量
  • PDF下载量:  0
  • 文章访问数:  624
  • HTML全文浏览量:  4
文章相关
  • 收稿日期:  2011-03-10
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

/

返回文章