Extensible and swellable hydrogel-forming microneedles for deep point-of-care sampling and drug deployment

Yuan Liu Ting Huang Zhiyong Qian Wei Chen

Citation:  Yuan Liu, Ting Huang, Zhiyong Qian, Wei Chen. Extensible and swellable hydrogel-forming microneedles for deep point-of-care sampling and drug deployment[J]. Chinese Chemical Letters, 2023, 34(6): 108103. doi: 10.1016/j.cclet.2022.108103 shu

Extensible and swellable hydrogel-forming microneedles for deep point-of-care sampling and drug deployment

English

  • Microneedles (MNs) are a promising and innovative transdermal drug delivery system with excellent permeability and functional diversity [1]. According to the materials and structures, MNs are mainly divided into 4 categories (solid, coated, hollow and dissolving MNs), which are extensively applied in the fields of tumor therapy, vaccine delivery, biomarker detection, and biosensing [2-4]. However, the solid MN delivery involves a two-step process known as the "poke and patch" method, which dramatically reduces patient compliance [5]. Besides, the direct deposition of the drug to the surface of the coated MNs significantly reduces the mechanical strength and sharpness of the needle, limiting the drug load to a low level [6-8]. In addition, the hollow MNs are prone to undergo blockage by skin tissue and spillage of the drug out of injecting site [4,9-12]. Notably, the main components of those MNs, such as silicon, glass, ceramics, metal, and stainless steel, exhibited relatively low biocompatibility with a high fracture rate, which brings potential risks to the human body and limits their further development [13-16]. Although dissolving MNs with drugs encapsulated in a water-soluble polymer matrix is able to reduce the risk, the potential for partial deposition of the polymer in the skin after dissolution in vivo to form granules or erythema is also a non-negligible problem [17,18].

    To overcome the limitations of traditional MNs design, extensible and swellable hydrogel-forming microneedles (ES-HFMs) were proposed, which rely on the deformation effect of hydrogels to produce a continuous, non-blocking pathway from the drug reservoir to the microcirculation within the dermis [19-21]. Consisting of cross-linked hydrogel polymers with dynamic deformation properties based on polymer chain relax, ES-HFMs feature a distinct operating mechanism to other conventional MNs (Fig. 1). The hydrophilic nature of the hydrogel polymer allows the ES-HFMs to swell and extend upon insertion into the skin, imparting a deep interaction between the system and local tissues. Alternatively, ES-HFMs offer the possibility of smart drug deployment, by regulating the ratio of polymeric crosslinkers to achieve controlled drug release [22-26]. Importantly, the hydrogel fabrication is flexibly amendable, as the system with various mechanical properties could be adjusted by formulation alteration and dimension restriction [27-31]. Furthermore, the hydrogel systems surmount the biocompatibility problems of silicon or metal MNs while allowing for completed removal from the skin or full dissociation with significant biosafety [32-36]. Evidently, ES-HFMs hold enormous scope for biomedical applications, as in minimally invasive methods of extracting interstitial skin fluid (ISF) for biomarker detection. In this review, we systematically summarized the recent development of ES-HFMs based on advances in the fields of chemistry, medicine and engineering. Initially, the design and fabrication strategy of ES-HFMs was described in detail, especially in terms of bionics and engineering inspiration. Subsequently, the application of the ES-HFMs was depicted with emphasis on the potential in deep point-of-care testing and drug deployment. Furthermore, the challenges and prospects, including biosecurity and clinical translational potential, were discussed in depth. It is believed that the ES-HFMs will be a promising candidate in industrial and clinical fields, indicating great potential in healthcare providers and patients in the near future.

    Figure 1

    Figure 1.  Schematic representation of the dynamic deformation of ES-HFMs. (A) Rapid extension and swelling of the compressed network of ES-HFMs when exposed to water, demonstrating excellent dynamic shape alteration capabilities. (B) Typical ES-HFMs applications including deep point-of-care sampling, smart drug deployment, wound healing and mechanical interlocking are illustrated.

    During the ES-HFMs fabrication, diverse polymers serve as crucial components as a result of their gelling capability, which effectively improves the mechanical strength of the system and facilitates controlled drug release, as well as dynamic shape alteration [29,37-39]. The materials include natural polymers (e.g., hyaluronic acid, gelatin, agar, silk protein) and synthetic polymers (e.g., polyvinyl alcohol, sodium polyacrylate, poly(methacrylic acid), acrylic resin, polyvinyl pyrrolidone), can generate a broad spectrum of hydrogels by a physical connection, electrostatic interaction or chemical cross-linking between molecules [40]. For example, Guozhong Yang et al. utilized acrylic resin to develop ES-HFMs with satisfactory mechanical properties and drug loading capacity, which greatly improved the release of water-insoluble drug, granisetron base [41]. Further studies revealed that the system obtained a 7-day controlled release of granisetron base when involving porogenic agents (1.5% dicalcium phosphate and 1.5% polyvinyl pyrrolidone). Meanwhile, the maximum drug amount entering the skin was 86.158% ± 7.82% of the initial granisetron base (2.1 mg), suggesting that systems incorporating appropriate polymeric materials offer a minimally invasive route but deep penetration and amount of drug deployment. Interestingly, it was found that certain polymers are able to smartly respond to the microenvironmental stimuli including pH, light or temperature, exhibiting significant dimension changes through swelling, expansion, sol-gel transition and network re-structuring. For instance, John G. Hardy et al. prepared stimulus-responsive HFMs using 2-hydroxyethyl methacrylate and ethylene glycol dimethacrylate, which are capable of delivering clinically used model drugs (ibuprofen) under light stimulation through spontaneous swelling (Fig. 2) [42]. Notably, this type of material was found to be more suitable for long-acting drug delivery due to the controllable and adjustable swelling processes.

    Figure 2

    Figure 2.  Materials for ES-HFMs. (A) Schematic representation of drug release from ES-HFMs made from light-responsive materials. (B) Cumulative fractional release profiles of ibuprofen from 3 × 3 ES-HFMs loaded with photoresponsive couplers over 3 irradiation cycles (irradiation 1/2/3) after 1 h of continuous irradiation at 365 nm. (C) Ibuprofen release from 3 × 3 ES-HFMs with release triggered by 3 cycles (dose 1/2/3) of exposure to light for 1 h. The 1% and 5% loading are ES-HFMs containing 1% and 5% light-responsive couplers. Copied with permission [42]. Copyright 2016, the American Chemical Society Publications.

    Due to the heterogeneity of the skin and unavoidable movement during puncture, ES-HFMs are subject to a variety of mechanical interactions, leading to fracture, bending, and other deformations, which affect the puncture performance [26,35,43,44]. Therefore, besides the matrix material, the engineering structure of the MNs such as length, shape, density, and spacing distance between each needle play a significant role in the penetration and removal processes [45-47]. The length of the MNs needs to be optimized to guarantee that the needles penetrate the skin whilst avoiding touching the underlying nerve endings, hence most MNs are approximately 25 µm to 2500 µm in length, 50 µm to 250 µm in width, and 1 µm to 25 µm in tip diameter [48]. Over-dense MNs would result in a "bed of nails" effect, which makes it difficult to effectively break through the stratum corneum barrier. At the same time, it decreases the MNs spacing, which remarkably reduces the penetration rate especially when the spacing is below 150 µm. According to a recent study, a density of fewer than 2000 needles/cm2 would be more favorable in terms of effective skin penetration [49,50]. For example, Zequan Zhou et al. constructed a Ca2+ cross-linked alginate ES-HFMs (round, 500 µm height, 500 µm spacing, 145 tips), which significantly improved the drug transdermal penetration, especially for acidic drugs [51]. Noticeably, differences in matrix materials, geometry parameters, and applicable targets may lead to variations in results, a further systematic exploration is still necessary to study the relationship between engineering structure and ES-HFMs performance to provide sufficient controllability, effectiveness and safety.

    Recently, bionics has gained substantial interest in the ES-HFMs fabrication, which inspired scientists and engineers to optimize the structure and shape layout of MNs by simulating the unique physiological and functional properties of living organisms [52,53]. Based on the biological surface wetting, adhesion, and friction mechanisms, the combination of the mechanics, materials science, mechanical design and manufacturing science, achieves great success in surface wetting, adhesion adjustment, resistance enhancement, and wear tolerance [54]. The stinging-sucking mouthparts of mosquitoes, gadflies, bees, cicadas and other insects have been extensively studied and serve as bionic prototypes of MNs due to their promising epidermis penetration and smart interactions [55-57]. Interestingly, in nature, some internal parasites have evolved to adhere to the host's intestinal wall by utilizing special components (e.g., hooks or suckers). Pomphorhynchus laevis, known as the spiny-headed worm, uses an expandable proboscis to hold firmly in the parasitic position after penetrating the intestinal wall [58]. Utilizing this principle, Eun Young Jeon et al. designed a biomimetic swellable hydrogel bilayer MN that achieves excellent tissue adhesion and closure under wet and/or dynamic conditions through surface adhesion and selective swellability of the hydrogel (Fig. 3) [32].

    Figure 3

    Figure 3.  Engineering structure of ES-HFMs. (A) Schematic diagram of the working mechanism of ES-HFMs, a patch consisting of a swellable and adhesive mussel adhesive proteins (MAP)-based shell and a non-swellable silk fibroin (SF)-based core. (B) Macroscopic images of rat skin wound healing after non-treatment, suture treatment and 60% double-layered MN (DL-MN) treatment. (C) Macroscopic images of a 5 mm diameter defect in the rat ileum after untreated, suture-treated and mussel adhesive protein/hyaluronic acid single-layered MN (MAP/HA SL-MN) treatment. Defect (black star), treated site (black dashed circle), the border between attached bowel (yellow dashed line), adhesion site indicating intestinal canal (yellow arrow), fatty attachment (black arrow) and leakage (yellow triangle). Copied with permission [32]. Copyright 2019, the Elsevier.

    MN-mediated drug release generally relies on passive diffusion, where external stimuli (e.g., ultrasound, light, temperature) are often employed to improve drug delivery, avoiding the low penetration and uneven distribution of therapeutic agents [59,60]. For example, Lu Fan et al. proposed a strategy for versatile and controlled drug delivery by integrating photoresponsive drug delivery microspheres into pyramidal MNs [61]. Benefiting from the sufficient water content of the swellable hydrogel, the MN was able to load various bioactive substances with well-maintained biological activity, which could be delivered to a deeper location at the target site. Nevertheless, the demand of bulky external equipment limits its application as a result of the troublesome operation, specialist handling and high cost. Further efforts are thus necessary to incorporate the merits of autonomous and active drug delivery into a simple MN delivery platform, while decreasing the time and cost required to achieve significant therapeutic efficiency. Recently, nano-scaled components were encapsulated into the ES-HFMs to endow them with autonomous bioactive properties, facilitating the MN extension and drug diffusion. As an illustration, Miguel Angel Lopez-Ramirez et al. described a swellable hydrogel-based bioactive MN delivery platform loaded with a therapeutic payload and active motor magnesium (Mg) microparticles, which was capable of generating autonomous intense convective fluid flow to improve drug deployment (Figs. 4A-D) [62]. As the polymer matrix dissolves within the skin, the embedded magnesium particles were exposed and react instantly with the surrounding ISF to rapidly produce hydrogen bubbles (Fig. 4E). These microbubbles induce a powerful and autonomous swelling pumping effect, and thus result in dynamically and extremely efficient transport and penetration of the embedded therapeutic payload. In vitro evaluation using fluorescence dye as a model drug revealed that the encapsulated Mg particles greatly enhanced the effective loading and delivery of the model dye within 5 min compared to the delayed delivery of passive MNs (without Mg particles). This phenomenon was significantly more pronounced at 10 min, where the passive MN resulted in a mean depth of 111 ± 79 µm versus 526 ± 71 µm with the active MN (Fig. 4F). Distinctively, the active delivery strategy of ES-HFMs offers considerable promise for rapid and deep drug deployment with a sophisticated design and bioactive nanocomponent integration.

    Figure 4

    Figure 4.  Bioactive nanocomponents of ES-HFMs. (A) The active MN patch component, as well as the built-in Mg particles that act as a pump when in contact with body fluids, are activated, thus enhancing drug release. (B) Digital photograph showing an optical/fluorescent microscope image of a 15 × 15 MN array and an active MN tip. (C) SEM image of a single active MN tip and EDX analysis of Mg. (D) Manufacturing steps for MNs: poly(dimethylsiloxane) microforming on the primary MN, poly(dimethylsiloxane) negative MN demolding, Mg particle loading, polymer and payload inclusion, polymer drying, adhesive application and peeling. (E) Microscope time frame image is taken from a single reactive MN point, clearly showing the dissolution of the polymer in PBS pH 6.0 and the activation of the particles (at 30-second intervals). (F) Schematic diagram illustrating the experimental set-up of passive (ⅰ) and active (ⅱ) MNs penetrating pig skin, and cross-sectional images of fluorescence microscopy at different times. Copied with permission [62]. Copyright 2020, the Wiley.

    Nowadays, the HFMs have been extensively studied and exhibited great potential in deep point-of-care sampling, smart drug deployment, wound healing, mechanical interlocking, etc. (Table 1). Currently, the most widespread approach for ES-HFMs fabrication should be the micro-modeling, in which the polymer solution was filled into the prepared microneedle array mold by external forces such as heating, centrifugation, evacuation, and light-mediated crosslinking to form the microneedle array (Fig. 5). As an illustration, Na Xu et al. prepared swellable polyvinyl alcohol/polyvinylpyrrolidone hydrogel microneedle patches by casting the polymer solution into the polydimethylsiloxane (PDMS) molds followed by careful removal, which perfectly maintained the morphology and were able to penetrate the skin in the dry state to extract ISF [26]. Furthermore, in order to minimize tissue damage while achieving significant adhesion to soft tissues, Yun Yang et al. applied a similar method to create a stimulus-responsive bilayer microneedle via double casings of block polyacrylic acid (PS-b-PAA) block copolymer and polystyrene (PS) [58]. Briefly, when PS-b-PAA solution was added into a PDMS mold, capillary forces would induce a thick film near the conical cavity's tip, while a thin film was generated elsewhere on the mold surface. Subsequently, the PS homopolymer was then melted on top of the intact PS-b-PAA layer, and the microneedle patch was peeled from the PDMS after cooling to the room temperature.

    Table 1

    Table 1.  Typical ES-HFMs and their applications.
    DownLoad: CSV

    Figure 5

    Figure 5.  The typical manufacturing process of ES-HFMS via micro-modeling.

    Interstitial skin fluid (ISF) containing a wealth of biological information is an ideal sample resource for point-of-care testing; however, collecting sufficient markers in ISF in a minimally invasive or non-invasive manner remains a great challenge [63,64]. In recent years, solid and hollow MNs have attracted growing attention in ISF sampling by facilitating fluid transport due to the capillary action [65]. Nevertheless, these systems are mainly made of silicon, glass and metal, which suffer from fragility, environmental pollution, complicated preparation process, high cost and short skin micropore opening time. Moreover, in order to speed up the extraction process and obtain sufficient ISF, multiple steps and additional devices to provide negative pressure are often required, severely increasing the complexity and risk of infection for the user. Therefore, it is particularly important to continuedly develop user-friendly and commercially MNs for deep point-of-care testing.

    The specific phase change properties of hydrogels (stiff in drying, soft after absorbing water) result in accessible skin penetration, extensible and swellable capabilities, perfectly suitable for non-pump ISF extraction [66]. Rongyan He et al. developed a point-of-care testing platform based on polyvinyl alcohol/chitosan hydrogel MNs, which allowed for quick extraction of rabbit skin ISF, the release of target biomarkers and analysis of their concentrations by colorimetric assay (darker color indicated higher glucose concentration) in a short time without any bulky instrumentation (Figs. 6A-C) [67]. It was observed that the ISF extracted from rabbits by this system increased with the extraction time and reached 1.25 ± 0.37 mg after 10 min (Fig. 6D). Meanwhile, the accuracy of ES-HFMs-based test was compared to hematological testing in terms of blood glucose level (BGL) which was performed using a commercial glucose meter, indicating excellent consistency between the measurements with the similar BGL range (4-14 mmol/L) (Fig. 6E). This phenomenon has also been verified in other biomarker detection (lactic acid, urea, etc.), suggesting that the ES-HFMs platform was capable of enhancing the point-of-care sampling with satisfactory accuracy, and without undesirable drawbacks of blood-related sampling, including vascular damage, bacterial infections, painful sensations, needle fear and self-administration challenges [40]. Afterward, insertion of the ES-HFMs into the skin of the rabbits revealed no breakage despite a slightly blunted tip, and the micropores disappeared rapidly within 3 min with minimal mechanical damage to the skin (Figs. 6F and G). Not negligibly, certain ES-HFMs still demand a multi-step sequence with high time consumption to gather sufficient volume (1 mL) for valid analysis. In order to simplify the process, Mengjia Zheng et al. introduced a swellable hydrogel (methacrylated hyaluronic acid) MN patch powered by an osmolyte (maltose), where the osmolyte dissolved in the matrix and provided osmotic pressure, increasing the diffusion of ISF from the skin to the hydrogel matrix [36]. The system with 100 MNs extracted 7.90 µL of ISF from in vitro pig skin and 3.82 µL of ISF from in vivo mouse skin within 3 min, whereas the control (hydrogel MNs free of penetrant) required more than 10 min to reach comparable outcomes. Given the latest trends in effortless personal self-medical monitoring, it would be more convenient to involve the biomarker enrichment, collection, release and analysis in one system. Developing automated ES-HFM-based platforms for sample concentration, microfluidics, and quick indication, or combining with other smart devices would be an encouraging strategy in the future point-of-care sampling.

    Figure 6

    Figure 6.  Application of ES-HFMs for deep point-of-care sampling. (A) Preparation of PVA/CS hydrogel ES-HFMs. (B) Target biomarkers were extracted from simulated skin in vitro and rabbit dorsal skin in vivo. (C) Recovery of target biomarkers from ES-HFMs by heating and detection using a colorimetric assay (darker color indicated higher glucose concentration). (D) The mass of ISF extracted from rabbits by the patch increased with extraction time and reached 1.25 ± 0.37 mg after 10 min. (E) Glucose levels obtained by ES-HFMs and a commercial glucose meter were excellent consistent in terms of quality and quantity. (F) Images before and after insertion of ES-HFMs into the skin of rabbits showing no breakage despite a slightly blunted needle tip. (G) Microchannels on the skin surface of the rabbit's back disappeared rapidly within 3 min after removal of ES-HFMs. Copied with permission [67]. Copyright 2020, the Wiley.

    The extensive property of the ES-HFMs provide enlarged space for drug deployment, which has been successfully employed to improve transdermal delivery of small molecule drugs such as granisetron base, sodium ibuprofen and lidocaine hydrochloride. Ryan F. Donnelly et al. developed "super-swelling" HFM containing 20% w/w Gantrez S-97, 7.5% w/w PEG 10,000, and 3% w/w Na2CO3, for the delivery of ibuprofen sodium assisted by the lyophilized wafer technology (Figs. 7A-D) [68]. These MN-based freeze-dried wafer compositions solidly and effectively penetrated the skin and swelled remarkably toward deep locations. During in vivo test, the system delivered approximately 44 mg of ibuprofen sodium within 24 h, equivalent to 37% of the load in the lyophilized storage solution. Besides, macromolecules were also deployed by this system, where 1.24 mg of ovalbumin was successfully delivered within one day, demonstrating a efficiency of about 49% (Figs. 7E-H). Furthermore, ES-HFMs with sustained drug release had also been invented, where swellable aqueous silk fibroin was fabricated in system with tremendous fidelity and mechanical robustness, allowing easy puncture into pig skin (depth of about 200 µm, and conversion to 50-700 nmol/L semisolid hydrogels with an internal porous network), which exhibited the sustained glucose release due to its gradual degradation [31]. Moreover, Siyuan Chen et al. reported a smart MN composed of silk protein and a semi-interpenetrating network hydrogel prepared from phenylboronic acid/acrylamide for glucose-responsive insulin delivery [69]. This smart MN autonomously releases insulin by modulating the skin layer formed on the surface, which corresponds favorably to the glucose variation pattern and promises to deliver on-demand insulin in a long-lasting, painless and convenient manner. It should be noted that nanoparticles (liposomes, microspheres, etc.) were frequently employed for subcutaneous drug delivery, but they were mainly compromised by the injection-associated pain, unremovable operation, uncontrollable degradation and limited penetration depth, which were fortunately impro ved by the ES-HFMs systems by offering smart insertion, swell and extension-based features [48,70-72].

    Figure 7

    Figure 7.  Application of ES-HFMs for smart drug deployment. (A) Schematic diagram of the application and retention strategy for rat experiments designed to assess the in vivo performance of ES-HFMs. (B) After 24 h of insertion, despite extensive swelling, ES-HFMs can still be removed intact from rat skin. (C) Digital image of a lyophilized wafer loaded with sodium ibuprofen. (D) Digital image of lyophilized wafer loaded with OVA. The plasma concentration of ibuprofen sodium (E) and OVA (F) in rats after transdermal administration using ES-HFMs. (G) Cumulative permeation profile of ibuprofen sodium in porcine skin by in vivo ES-HFMs combined with a lyophilized drug reservoir. (H) Cumulative permeation profile of OVA in porcine skin by in vivo ES-HFMs combined with a lyophilized drug reservoir. Copied with permission [68]. Copyright 2014, the Plos One.

    To repair severe tissue damage and skin scars, patches that significantly inhibit bacterial infection and accelerate wound healing are considered to be the most effective candidates [73-78]. Typically, the traditional formulation mainly included various active drugs (e.g., bactericidal, hemostatic, anti-inflammatory, and angiogenic factors) with adjuvants such as diverse nanoparticles [79-82]. Nonetheless, the complex biological and chemical fabrication procedures involved in conventional patches reduce biocompatibility and biosafety, thus further limiting their application in deep tissue injuries. Crucially, the single structure does not confer extensible and swellable dynamic shape alteration properties to the typical patches, and thus fails to provide real-time traction and fixation effects for tissue repair. As a result, it is expected to develop multifunctional patches for wound healing in combination with the tunability of hydrogel interactions and the superior microstructure of MNs [83-88]. Junjie Chi et al. invented a biomass chitosan-ES-HFMs by utilizing poly(N-isopropylacrylamide) hydrogel as an intelligent temperature responsive system, which was able to encapsulate vascular endothelial growth factor (VEGF) in the micropores [89]. Precisely controlled release of VEGF was achieved through exposure to different temperatures that cause the poly(N-isopropylacrylamide) hydrogel to extend/contract, thus accelerating inflammation inhibition, collagen deposition, angiogenesis and granulation tissue formation. Furthermore, Soomee Lim et al. proposed a double-layered adhesive MN bandage (DL-AMNB) as a therapeutic cardiac patch, which consisted of a biofunctional mussel adhesive protein-based root and a regenerating filamentous protein-based tip (Fig. 8) [90]. The system evidently displayed better cardiac muscle preservation and wound healing compared to the conventional cardiac patches in a rat myocardial infarction model, possibly owing to highly efficient delivery and long-term retention of the therapeutic peptide (i.e., vascular endothelial growth factor, fibroblast growth factor-2 and extracellular matrix protein), as well as the safe adhesion between the patch and the host myocardium via the intense underwater adhesion similar to mussel adhesive protein.

    Figure 8

    Figure 8.  Application of ES-HFMs for wound healing. (A) Schematic representation of DL-AMNB, including biofunctional peptides fusion mussel adhesive protein for myocardial infarction treatment. (B) In vivo DL-AMNB treatment effect in a rat myocardial infarction model. Photographs of DL-AMNB treatment of the infarct zone before (ⅰ) and after (ⅱ). The yellow dashed circles represent the infarcted area. (ⅲ) Macroscopic images of the remaining DL-AMNB adhering to heart tissue 2 weeks after treatment. Copied with permission [90]. Copyright 2021, the Elsevier.

    Achieving mechanical interlocking effects on soft tissues while minimizing tissue damage is a considerably clinical challenge. Chemical adhesive-based approach requires specific reactive chemistry and may induce a severe inflammatory response [91-95]. Sutures need sophisticated operation by the professional staff, and are not suitable in dura mater, urethral defects, lung tissue and so on [96-100]. Staples are also fraught with limitations including intense local tissue pressure and high infection risk, as well as nerve and vascular damage [101-105]. Unlike ordinary mechanical interlocking means, ES-HFMs have displayed shorter procedure times, decreased scarring rate, dispersed force over a broader surface area, as well as dynamically scalable tissue handling.

    Inspired by the swollen proboscis of endoparasites for stable immobilization on host tissue after insertion, Seung Yun Yang et al. have developed a biphasic MN that exhibited precise interlocks with the tissue through an expandable MN tip, further developing grip strength by about 3.5 times contrasted with staples in the skin for uniting obsession with an evacuation force about 4.5 N/cm2 from gastrointestinal mucosal tissue (Fig. 9) [58]. This design minimizes the force required to penetrate the tissue, ensuring firm fixation while allowing removal without significant damage to the tissue. In addition, dissimilar to unbending MNs, the powerful dynamic shape alteration properties of ES-HFMs keep them from bursting when eliminated from the tissue. Furthermore, Eun Young Jeon et al. designed mutually cross-linked biphasic ES-HFMs, involving a water-absorbable adhesive exterior based on swellable mussel adhesive protein and a rigid interior consisting of non-swellable silk fibroin-based core [32]. With excellent tissue insertion, surface adhesion and mechanical interlocking ability, the device achieves superior wound sealing capacity in vitro (139.7 ± 14.1 mmHg), comparable to clinically used sutures (151.0 ± 23.3 mmHg), offering immediate closure of bleeding intestinal wounds in vivo without leakage, abscesses or bowel obstruction.

    Figure 9

    Figure 9.  Application of ES-HFMs for mechanical interlocking. (A) The mechanical interlocking of ES-HFMs after tissue penetration. (B) ES-HFMs were prepared using PDMS molds and show the chemical structure of polystyrene-block-poly-acrylic acid (PS-b-PAA), where the weight fraction of polystyrene used for the expandable tip is 25%. (C) Schematic diagram showing the internal structure and reversible water reactivity of a bilayer MN. (D) Side views of hollow MNs (without polystyrene core) with different PS-b-PAA filling ratios, including (ⅰ) 20%, (ⅱ) 40% and (ⅲ) 70%. (ⅳ) Bilayer MN with expandable tip after filling with polystyrene core. (E) Photograph of ES-HFMs at a density of 10 × 10 cm−2, showing high pattern fidelity. Copied with permission [58]. Copyright 2013, the Nature publishing group.

    Interestingly, Wei Chen et al. described an intelligent targeted capsule system that combines physical (MN) and non-physical (enhancer) drug delivery modes [106]. To build smart MNs with extension and tissue penetration, the dynamic omnidirectional adhesive MN system (DOAMS) featuring a biphasic core-shell structure, a soft outer layer (Carbopol) and a rigid inner core (polycaprolactone), was designed inspired by the natural thorny-headed intestinal worm (Fig. 10A). Afterward, a self-triggering capsule was designed inspired by the "jack-in-the-box" system, which delivers a tablet loaded with MNs to the stomach, and then ejects it into the stomach cavity when triggered by the local gastric environment (Fig. 10B). DOAMS showed extremely high adhesion to in vitro pig stomach tissue (0.25 N), even stronger than commercial Carbopol powders (0.18 N). In vivo, real-time endoscopic imaging showed that conventional tablets (non-mucosal adhesives and non-MNs) migrated more than 30 mm in 20 s, while DOAMS-modified tablets stayed securely in place (Figs. 10C-E). This might be attributed to the presence of multiple interlocking mechanisms among DOAMS and the tissue: mechanical interlocking driven hooks and chemical adhesion of the hydrogel bonds between Carbopol and the tissue.

    Figure 10

    Figure 10.  Application of ES-HFMs for mechanical interlocking based on thorny-headed worm. (A) (ⅰ) Thorny-headed worm attaches to the intestine of fish. (ⅱ) The micro-needle-like structures in the head of the worm expand as they penetrate the intestine, providing strong adhesion. (ⅲ) After intake, the "jack-in-the-box" system would automatically deploy into the tissues in reaction to the low pH environment in the stomach. (B) Schematic diagram of a device comprising a double DOAMS modified tablet. (C) The image display device from the high-speed camera is successfully triggered. (D) Pictures of the state of the double DOAMS modified tablets before and after initiation. (E) Endoscopic images showing the movement of control and individual DOAMS-modified tablets in the pig stomach. Copied with permission [106]. Copyright 2022, the Science Advances.

    The extraordinary dynamic shape variability and interaction tunability of ES-HFMs address a critical headway over customary MNs frameworks, enormously broadening the scope of MNs applications. Nevertheless, still lots of issues remain to be clarified for further clinical translation and commercial application of ES-HFMs. (1) The variability among individual patient groups, (e.g., differences of water content in the skin of children, the elderly and pregnant women), poses a harsh challenge to precisely modulate the expansion coefficient and acting duration of ES-HFMs for personalized treatment. (2) Although the hydrogel polymers are produced with adequate cross-linking extent, undesired degradation may still occur during long-term action in vivo, thus worsening the inflammatory response in wet wounds. (3) MNs penetrating the stratum corneum are expected to avoid the introduction of harmful microorganisms into the skin, which frequently triggers an inflammatory response [107-111]. Different from the metals and silica MNs that use autoclaving process, ES-HFMs are typically sterilized using ultraviolet light irradiation, which presents tremendous challenges for light-sensitive or protein-based macromolecule encapsulation [112-116]. (4) In cases where ES-HFMs are not applied immediately after fabrication, the bioactive drug molecules on the surface may decrease activity during room temperature storage [117-119]. (5) Owing to being in small-scale research in experimental animal models, ES-HFMs needed to be tested in large animal models or humans in the near future [120-122].

    As MNs continue to progress in commercialization, there is a greater necessity to address the issues related to their translation from the lab to the market. (1) Predictably, the ES-HFMs are expected to act as a portable and integrated device, indicating a great potential to be applied to scenarios outside of the laboratory, such as home and hospitals. As an illustration, combining ES-HFMs with other cutting-edge technologies (e.g., microactuators, biosensors and microfluidic chips) to create intelligent microsystems to integrate sample collection/analysis and drug deployment [123-128]. (2) It is especially vital to enrich the dynamic shape alteration properties of ES-HFMs in fitting them better to implant sites containing irregular geometric shapes. Shimon Unterman et al. designed a nanocomposite hydrogel with independently tunable rheological and mechanical properties, consisting of dextran aldehyde and poly(amido amine) dendrimer mixed with phyllosilicate nanoparticles [129]. It has been proven that these nanocomposite hydrogels with different aspect ratios in combination with MNs could be optimally implanted in wet, fluid and inverted environments. (3) The optimal ES-HFM shapes are required to satisfy the needs of different individuals because the penetration of ES-HFMs varies with age, skin condition, and the applied force during the treatment. (4) The majority of currently available microneedle products are more expensive than conventional blood-based sample extraction and subcutaneous formulation injection; therefore, decreasing the cost of ES-HFMs and achieving mass production are crucial steps forward in the industry. Despite existing difficulties, ES-HFMs have been verified to exhibit considerable potential for deep point-of-care sampling and smart drug deployment by overcoming the shortcomings of conventional MNs. It is undoubtedly that ES-HFMs will continuedly serve as a promising candidate for smart biomedical applications in the clinical and industry in the future.

    The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

    This study was supported by the Fundamental Research Funds for the Central Universities (No. 5003510106), the National Natural Science Foundation of China (Nos. U21A20417, 31930067), and 1·3·5 Project for Disciplines of Excellence, West China Hospital, Sichuan University (No. ZYGD18002).


    1. [1]

      I. Menon, P. Bagwe, K.B. Gomes, et al., Micromachines (Basel) 12 (2021) 435. doi: 10.3390/mi12040435

    2. [2]

      Y. Hao, W. Li, X. Zhou, et al., J. Biomed. Nanotechnol. 13 (2017) 1581–1597. doi: 10.1166/jbn.2017.2474

    3. [3]

      T. Waghule, G. Singhvi, S.K. Dubey, et al., Biomed. Pharmacother. 109 (2019) 1249–1258. doi: 10.1016/j.biopha.2018.10.078

    4. [4]

      Á. Cárcamo-Martínez, B. Mallon, J. Domínguez-Robles, et al., Int. J. Pharm. 599 (2021) 120455. doi: 10.1016/j.ijpharm.2021.120455

    5. [5]

      Y. Mizuno, K. Takasawa, T. Hanada, et al., Biomed. Microdevices 23 (2021) 38. doi: 10.1007/s10544-021-00576-x

    6. [6]

      J. Xu, D. Xu, X. Xuan, H. He, Molecules 26 (2021) 5912. doi: 10.3390/molecules26195912

    7. [7]

      R. Jamaledin, C. Di Natale, V. Onesto, et al., J. Clin. Med. 9 (2020) 542. doi: 10.3390/jcm9020542

    8. [8]

      R.S.J. Ingrole, H.S. Gill, J. Pharmacol. Exp. Ther. 370 (2019) 555–569. doi: 10.1124/jpet.119.258707

    9. [9]

      P. Dardano, S. De Martino, M. Battisti, et al., Polymers (Basel) 13 (2021) 520. doi: 10.3390/polym13040520

    10. [10]

      V. Yadav, P.K. Sharma, U.S. Murty, et al., Int. J. Pharm. 605 (2021) 120815. doi: 10.1016/j.ijpharm.2021.120815

    11. [11]

      N.D. Sheybani, H. Yang, Chin. Chem. Lett. 28 (2017) 1817–1821. doi: 10.1016/j.cclet.2017.07.022

    12. [12]

      Y. Chen, X. Zhang, F. Liu, et al., Chin. Chem. Lett. 32 (2021) 441–444. doi: 10.1016/j.cclet.2020.05.009

    13. [13]

      M. Avcil, A. Çelik, Micromachines (Basel) 12 (2021) 1321. doi: 10.3390/mi12111321

    14. [14]

      D. Kulkarni, F. Damiri, S. Rojekar, et al., Pharmaceutics 14 (2022) 1097. doi: 10.3390/pharmaceutics14051097

    15. [15]

      J.H. Jung, S.G. Jin, J. Pharm. Investig. 51 (2021) 503–517. doi: 10.1007/s40005-021-00512-4

    16. [16]

      J. Yang, X. Liu, Y. Fu, Y. Song, Acta. Pharm. Sin. B 9 (2019) 469–483. doi: 10.1016/j.apsb.2019.03.007

    17. [17]

      R. Amarnani, P. Shende, Biomed. Microdevices 24 (2021) 4.

    18. [18]

      F.K. Aldawood, A. Andar, S. Desai, Polymers (Basel) 13 (2021) 2815. doi: 10.3390/polym13162815

    19. [19]

      R.F. Donnelly, T.R. Singh, M.J. Garland, et al., Adv. Funct. Mater. 22 (2012) 4879–4890. doi: 10.1002/adfm.201200864

    20. [20]

      Z. Xu, G. Liu, J. Huang, J. Wu, ACS Appl. Mater. Interfaces 14 (2022) 7680–7689. doi: 10.1021/acsami.1c23461

    21. [21]

      H. Chang, M. Zheng, X. Yu, et al., Adv. Mater. 29 (2017) 1702243. doi: 10.1002/adma.201702243

    22. [22]

      C.F. Guimarães, R. Ahmed, A.P. Marques, et al., Adv. Mater. 33 (2021) e2006582. doi: 10.1002/adma.202006582

    23. [23]

      S. Li, Y. Cong, J. Fu, J. Mater. Chem. B 9 (2021) 4423–4443. doi: 10.1039/D1TB00523E

    24. [24]

      A. Rajput, M. Kulkarni, P. Deshmukh, et al., Drug Dev. Ind. Pharm. 47 (2021) 1713–1732. doi: 10.1080/03639045.2022.2058531

    25. [25]

      K. Aich, T. Singh, S. Dang, Drug Deliv. Transl. Res. 12 (2022) 1556–1568. doi: 10.1007/s13346-021-01056-8

    26. [26]

      N. Xu, M. Zhang, W. Xu, et al., Analyst 147 (2022) 1478–1491. doi: 10.1039/D1AN02288A

    27. [27]

      R. Narayanaswamy, V.P. Torchilin, Molecules 24 (2019) 603. doi: 10.3390/molecules24030603

    28. [28]

      Z. Sun, C. Song, C. Wang, et al., Mol. Pharm. 17 (2020) 373–391.

    29. [29]

      L. Barnum, J. Quint, H. Derakhshandeh, et al., Adv. Healthc. Mater. 10 (2021) e2001922. doi: 10.1002/adhm.202001922

    30. [30]

      X. Chen, H. Yu, L. Wang, et al., ACS Biomater. Sci. Eng. 7 (2021) 4870–4882. doi: 10.1021/acsbiomaterials.1c01073

    31. [31]

      Z. Yin, D. Kuang, S. Wang, et al., Int. J. Biol. Macromol. 106 (2018) 48–56. doi: 10.1016/j.ijbiomac.2017.07.178

    32. [32]

      E.Y. Jeon, J. Lee, B.J. Kim, et al., Biomaterials 222 (2019) 119439. doi: 10.1016/j.biomaterials.2019.119439

    33. [33]

      Q. He, R. Kusumi, S. Kimura, et al., Carbohydr. Polym. 237 (2020) 116189. doi: 10.1016/j.carbpol.2020.116189

    34. [34]

      Y. Xu, H. Chen, Y. Fang, J. Wu, Adv. Health. Mater. 11 (2022) 2200494. doi: 10.1002/adhm.202200494

    35. [35]

      D.F.S. Fonseca, P.C. Costa, I.F. Almeida, et al., Macromol. Biosci. 20 (2020) e2000195. doi: 10.1002/mabi.202000195

    36. [36]

      M. Zheng, Z. Wang, H. Chang, et al., Adv. Healthc. Mater. 9 (2020) e1901683. doi: 10.1002/adhm.201901683

    37. [37]

      F. Meng, A. Hasan, M. Mahdi Nejadi Babadaei, et al., J. Adv. Res. 26 (2020) 137–147. doi: 10.1016/j.jare.2020.07.017

    38. [38]

      Y. Liu, D. Li, J. Ding, X. Chen, Chin. Chem. Lett. 31 (2020) 3001–3014. doi: 10.1016/j.cclet.2020.04.029

    39. [39]

      Y. Han, X. Li, Y. Zhang, et al., Cells 8 (2019) 886. doi: 10.3390/cells8080886

    40. [40]

      J.G. Turner, L.R. White, P. Estrela, H.S. Leese, Macromol. Biosci. 21 (2021) e2000307. doi: 10.1002/mabi.202000307

    41. [41]

      G. Yang, M. He, S. Zhang, et al., Drug Dev. Ind. Pharm. 44 (2018) 808–816. doi: 10.1080/03639045.2017.1414230

    42. [42]

      J.G. Hardy, E. Larrañeta, R.F. Donnelly, et al., Mol. Pharm. 13 (2016) 907–914. doi: 10.1021/acs.molpharmaceut.5b00807

    43. [43]

      Q.K. Anjani, A.D. Permana, Á. Cárcamo-Martínez, et al., Eur. J. Pharm. Biopharm. 158 (2021) 294–312. doi: 10.1016/j.ejpb.2020.12.003

    44. [44]

      K. Peng, L.K. Vora, J. Domínguez-Robles, et al., Mater. Sci. Eng. C: Mater. Biol. Appl. 127 (2021) 112226. doi: 10.1016/j.msec.2021.112226

    45. [45]

      T. Sheng, B. Luo, W. Zhang, et al., Adv. Drug Deliv. Rev. 179 (2021) 113919. doi: 10.1016/j.addr.2021.113919

    46. [46]

      X. He, J. Sun, J. Zhuang, et al., Dose Response 17 (2019) 1559325819878585.

    47. [47]

      P. Makvandi, M. Kirkby, A.R.J. Hutton, et al., Nanomicro. Lett. 13 (2021) 93.

    48. [48]

      K. Ahmed Saeed Al-Japairai, S. Mahmood, S. Hamed Almurisi, et al., Int. J. Pharm. 587 (2020) 119673. doi: 10.1016/j.ijpharm.2020.119673

    49. [49]

      I. Xenikakis, M. Tzimtzimis, K. Tsongas, et al., Eur. J. Pharm. Sci. 137 (2019) 104976. doi: 10.1016/j.ejps.2019.104976

    50. [50]

      K.J. Hsueh, M.C. Chen, L.T. Cheng, et al., Comp. Immunol. Microbiol. Infect. Dis. 50 (2017) 78–87. doi: 10.1016/j.cimid.2016.12.001

    51. [51]

      Z. Zhou, M. Xing, S. Zhang, et al., Int. J. Pharm. 618 (2022) 121669. doi: 10.1016/j.ijpharm.2022.121669

    52. [52]

      G. Ma, C. Wu, J. Control. Release. 251 (2017) 11–23. doi: 10.1016/j.jconrel.2017.02.011

    53. [53]

      M. Araújo, J. Silveira, A. Sousa, et al., Biomater. Sci. 9 (2021) 6510–6527. doi: 10.1039/D1BM00118C

    54. [54]

      H. Witte, Biomimetics (Basel) 7 (2022) 96. doi: 10.3390/biomimetics7030096

    55. [55]

      A.R. Dixon, I. Vondra, Materials (Basel) 15 (2022) 4587. doi: 10.3390/ma15134587

    56. [56]

      C. Plamadeala, S.R. Gosain, F. Hischen, et al., Biomed. Microdevices 22 (2019) 8.

    57. [57]

      K. Girija Sravani, R.K. Desala, P. Chand, et al., IEEE Trans. Nanobioscience 22 (2023) 237–244. doi: 10.1109/TNB.2022.3185633

    58. [58]

      S.Y. Yang, E.D. O'Cearbhaill, G.C. Sisk, et al., Nat. Commun. 4 (2013) 1702. doi: 10.1038/ncomms2715

    59. [59]

      A. Hou, B. Cohen, A. Haimovic, N. Elbuluk, Dermatol. Surg. 43 (2017) 321–339. doi: 10.1097/DSS.0000000000000924

    60. [60]

      S.H. Bariya, M.C. Gohel, T.A. Mehta, O.P. Sharma, J. Pharm. Pharmacol. 64 (2012) 11–29.

    61. [61]

      L. Fan, X. Zhang, X. Liu, et al., Adv. Healthc. Mater. 10 (2021) e2002249. doi: 10.1002/adhm.202002249

    62. [62]

      M.A. Lopez-Ramirez, F. Soto, C. Wang, et al., Adv. Mater. 32 (2020) e1905740. doi: 10.1002/adma.201905740

    63. [63]

      P.P. Samant, M.M. Niedzwiecki, N. Raviele, et al., Sci. Transl. Med. 12 (2020) eaaw0285. doi: 10.1126/scitranslmed.aaw0285

    64. [64]

      X. Zhang, G. Chen, F. Bian, et al., Adv. Mater. 31 (2019) e1902825. doi: 10.1002/adma.201902825

    65. [65]

      P.P. Samant, M.R. Prausnitz, Proc. Natl. Acad. Sci. U. S. A. 115 (2018) 4583–4588. doi: 10.1073/pnas.1716772115

    66. [66]

      R.F. Donnelly, K. Mooney, M.T. McCrudden, et al., J. Pharm. Sci. 103 (2014) 1478–1486. doi: 10.1002/jps.23921

    67. [67]

      R. He, Y. Niu, Z. Li, et al., Adv. Healthc. Mater. 9 (2020) e1901201. doi: 10.1002/adhm.201901201

    68. [68]

      R.F. Donnelly, M.T. McCrudden, A. Zaid Alkilani, et al., PLoS One 9 (2014) e111547. doi: 10.1371/journal.pone.0111547

    69. [69]

      S. Chen, H. Matsumoto, Y. Moro-Oka, et al., ACS Biomater. Sci. Eng. 5 (2019) 5781–5789. doi: 10.1021/acsbiomaterials.9b00532

    70. [70]

      Y. Lei, Y. Wang, J. Shen, et al., Sci. Adv. 8 (2022) eabl6449. doi: 10.1126/sciadv.abl6449

    71. [71]

      Y. Sun, Q. Zhou, Y. Du, et al., Small 18 (2022) e2201656. doi: 10.1002/smll.202201656

    72. [72]

      M.J. Ostro, P.R. Cullis, Am. J. Hosp. Pharm. 46 (1989) 1576–1587.

    73. [73]

      H. Chen, Y. Cheng, J. Tian, et al., Sci. Adv. 6 (2020) eaba4311. doi: 10.1126/sciadv.aba4311

    74. [74]

      J. Liu, M. Qu, C. Wang, et al., Small 18 (2022) e2106172. doi: 10.1002/smll.202106172

    75. [75]

      W. Lyu, Y. Ma, S. Chen, et al., Adv. Healthc. Mater. 10 (2021) e2100785. doi: 10.1002/adhm.202100785

    76. [76]

      R. Huang, X. Zhang, W. Li, et al., Adv. Sci. 8 (2021) e2100201. doi: 10.1002/advs.202100201

    77. [77]

      Y.S. Kim, D.K. Sung, W.H. Kong, et al., Biomater. Sci. 6 (2018) 1020–1030. doi: 10.1039/C8BM00079D

    78. [78]

      S. Orman, S. Yol, H. Uzun, et al., J. Invest. Surg. 33 (2020) 97–105. doi: 10.1080/08941939.2018.1473902

    79. [79]

      I. Woodhouse, S. Nejati, V. Selvamani, et al., ACS Appl. Bio. Mater. 4 (2021) 5405–5415. doi: 10.1021/acsabm.1c00087

    80. [80]

      V. Deineka, O. Sulaieva, N. Pernakov, et al., Mater. Sci. Eng. C: Mater. Biol. Appl. 120 (2021) 111740. doi: 10.1016/j.msec.2020.111740

    81. [81]

      K. Howk, J. Fortier, R. Poston, Ann. Vasc. Surg. 31 (2016) 186–195. doi: 10.1016/j.avsg.2015.09.007

    82. [82]

      S. García-Salinas, M. Evangelopoulos, E. Gámez-Herrera, et al., Int. J. Pharm. 577 (2020) 119067. doi: 10.1016/j.ijpharm.2020.119067

    83. [83]

      J. Kim, S.W. Kim, S. Park, et al., Adv. Healthc. Mater. 2 (2013) 1525–1531. doi: 10.1002/adhm.201200368

    84. [84]

      O. Hakimi, P.A. Mouthuy, A. Carr, Int. J. Exp. Pathol. 94 (2013) 287–292. doi: 10.1111/iep.12030

    85. [85]

      K.A. Zink, J.C. Mayberry, E.G. Peck, M.A. Schreiber, Am. Surg. 77 (2011) 438–442. doi: 10.1177/000313481107700419

    86. [86]

      N. R, K. M, P. J, et al., Colloids Surf. B: Biointerfaces 182 (2019) 110339. doi: 10.1016/j.colsurfb.2019.06.068

    87. [87]

      H.C. Tsai, G.R. Chang, H.C. Fan, et al., BMC Vet. Res. 15 (2019) 191. doi: 10.1186/s12917-019-1932-7

    88. [88]

      R. Li, K. Liu, X. Huang, et al., Adv. Sci. 9 (2022) e2105152. doi: 10.1002/advs.202105152

    89. [89]

      J. Chi, X. Zhang, C. Chen, et al., Bioact. Mater. 5 (2020) 253–259. doi: 10.1016/j.bioactmat.2020.02.004

    90. [90]

      S. Lim, T.Y. Park, E.Y. Jeon, et al., Biomaterials 278 (2021) 121171. doi: 10.1016/j.biomaterials.2021.121171

    91. [91]

      S.J. Marshall, S.C. Bayne, R. Baier, et al., Dent. Mater. 26 (2010) e11–e16.

    92. [92]

      J. Shin, E.H. Kang, S. Choi, et al., ACS Biomater. Sci. Eng. 7 (2021) 4230–4243. doi: 10.1021/acsbiomaterials.0c01414

    93. [93]

      C. Samarawickrama, A. Samanta, A. Liszka, et al., Cornea 37 (2018) 609–616. doi: 10.1097/ICO.0000000000001459

    94. [94]

      F.M. Kamer, J.H. Joseph, Arch. Otolaryngol. Head. Neck. Surg. 115 (1989) 193–197. doi: 10.1001/archotol.1989.01860260067016

    95. [95]

      C. Ma, J. Sun, B. Li, et al., Nat. Commun. 12 (2021) 3613. doi: 10.1038/s41467-021-23117-9

    96. [96]

      Y. Lin, Z. Chen, Y. Liu, et al., Drug Des. Devel. Ther. 16 (2022) 2707–2728. doi: 10.2147/DDDT.S375541

    97. [97]

      Y. Teng, X. Zhang, L. Da, et al., BMC Musculoskelet. Disord. 22 (2021) 571. doi: 10.1186/s12891-021-04457-5

    98. [98]

      O.E. Beidas, J.A. Gusenoff, Aesthet. Surg. J. 39 (2019) S85–s93. doi: 10.1093/asj/sjz001

    99. [99]

      M. Byrne, A. Aly, Aesthet. Surg. J. 39 (2019) S67–S72. doi: 10.1093/asj/sjz036

    100. [100]

      B. Krishnamoorthy, N. Shepherd, W.R. Critchley, et al., Interact. Cardiovasc. Thorac. Surg. 22 (2016) 161–167. doi: 10.1093/icvts/ivv314

    101. [101]

      A. Cromi, A.S. Laganà, F. Ghezzi, et al., Eur. J. Obstet. Gynecol. Reprod. Biol. 271 (2022) 112–116. doi: 10.1016/j.ejogrb.2022.02.009

    102. [102]

      B.J. Cole, E.T. Sayegh, A.B. Yanke, et al., J. Am. Acad. Orthop. Surg. 24 (2016) 83–95. doi: 10.5435/JAAOS-D-14-00081

    103. [103]

      C. Salyer, A. Spuzzillo, D. Wakefield, et al., J. Surg. Res. 267 (2021) 705–711. doi: 10.1016/j.jss.2021.06.024

    104. [104]

      R. Zan, H. Wang, J. Ni, et al., ACS Biomater. Sci. Eng. 7 (2021) 5269–5278. doi: 10.1021/acsbiomaterials.1c00683

    105. [105]

      E.J. Hyland, D. Maze, T. Lawrence, et al., Int. Wound J. 13 (2016) 878–879. doi: 10.1111/iwj.12398

    106. [106]

      W. Chen, J. Wainer, S.W. Ryoo, et al., Sci. Adv. 8 (2022) eabk1792. doi: 10.1126/sciadv.abk1792

    107. [107]

      R.K. Sivamani, D. Liepmann, H.I. Maibach, Expert. Opin. Drug Deliv. 4 (2007) 19–25. doi: 10.1517/17425247.4.1.19

    108. [108]

      D. Li, D. Hu, H. Xu, et al., Biomaterials 264 (2021) 120410. doi: 10.1016/j.biomaterials.2020.120410

    109. [109]

      T.T. Nguyen, Y. Oh, Y. Kim, et al., Hum. Vaccin. Immunother. 17 (2021) 316–327. doi: 10.1080/21645515.2020.1767997

    110. [110]

      N.S. Rejinold, J.H. Shin, H.Y. Seok, Y.C. Kim, Expert Opin. Drug Deliv. 13 (2016) 109–131. doi: 10.1517/17425247.2016.1115835

    111. [111]

      J. Richter-Johnson, P. Kumar, Y.E. Choonara, et al., Expert Rev. Pharm. Out. 18 (2018) 359–369.

    112. [112]

      H. Lu, S. Zada, L. Yang, H. Dong, Front. Bioeng. Biotechnol. 10 (2022) 851134. doi: 10.3389/fbioe.2022.851134

    113. [113]

      A.K. Gera, R.K. Burra, J. Funct. Biomater. 13 (2022) 81. doi: 10.3390/jfb13020081

    114. [114]

      P.R. Yadav, M.N. Munni, L. Campbell, et al., Pharmaceutics 13 (2021) 1132. doi: 10.3390/pharmaceutics13081132

    115. [115]

      F. Zhang, W. Bao, R. Li, et al., Biomater. Sci. 7 (2019) 4503–4507. doi: 10.1039/C9BM00482C

    116. [116]

      K. Ita, Pharmaceutics 7 (2015) 90–105. doi: 10.3390/pharmaceutics7030090

    117. [117]

      A. Sivaraman, A.K. Banga, Drug Deliv. Transl. Res. 7 (2017) 16–26. doi: 10.1007/s13346-016-0328-5

    118. [118]

      N.N. Aung, T. Ngawhirunpat, T. Rojanarata, et al., Int. J. Pharm. 586 (2020) 119508. doi: 10.1016/j.ijpharm.2020.119508

    119. [119]

      J.Y. Li, Y.H. Feng, Y.T. He, et al., Acta Biomater. 153 (2022) 308–319. doi: 10.1016/j.actbio.2022.08.061

    120. [120]

      X. Chen, L. Wang, H. Yu, et al., J. Control. Release 288 (2018) 173–188. doi: 10.1016/j.jconrel.2018.08.042

    121. [121]

      R.F. Donnelly, D.I. Morrow, M.T. McCrudden, et al., Photochem. Photobiol. 90 (2014) 641–647. doi: 10.1111/php.12209

    122. [122]

      N.G. Oh, S.Y. Hwang, Y.H. Na, ACS Omega 7 (2022) 25179–25185. doi: 10.1021/acsomega.2c01993

    123. [123]

      C.G. Li, C.Y. Lee, K. Lee, H. Jung, Biomed. Microdevices 15 (2013) 17–25. doi: 10.1007/s10544-012-9683-2

    124. [124]

      Z. Faraji Rad, P.D. Prewett, G.J. Davies, Beilstein. J. Nanotechnol. 12 (2021) 1034–1046. doi: 10.3762/bjnano.12.77

    125. [125]

      P.C. Pandey, G. Pandey, R.J. Narayan, J. Biomed. Mater. Res. B: Appl. Biomater. 109 (2021) 33–49. doi: 10.1002/jbm.b.34678

    126. [126]

      P. Bollella, S. Sharma, A.E.G. Cass, R. Antiochia, Biosens. Bioelectron. 123 (2019) 152–159. doi: 10.1016/j.bios.2018.08.010

    127. [127]

      M. Jeyhani, V. Gnyawali, N. Abbasi, et al., J. Colloid. Interface Sci. 553 (2019) 382–389. doi: 10.1016/j.jcis.2019.05.100

    128. [128]

      S.T. Sanjay, W. Zhou, M. Dou, et al., Adv. Drug Deliv. Rev. 128 (2018) 3–28. doi: 10.1016/j.addr.2017.09.013

    129. [129]

      S. Unterman, L.F. Charles, S.E. Strecker, et al., ACS Nano 11 (2017) 2598–2610. doi: 10.1021/acsnano.6b06730

  • Figure 1  Schematic representation of the dynamic deformation of ES-HFMs. (A) Rapid extension and swelling of the compressed network of ES-HFMs when exposed to water, demonstrating excellent dynamic shape alteration capabilities. (B) Typical ES-HFMs applications including deep point-of-care sampling, smart drug deployment, wound healing and mechanical interlocking are illustrated.

    Figure 2  Materials for ES-HFMs. (A) Schematic representation of drug release from ES-HFMs made from light-responsive materials. (B) Cumulative fractional release profiles of ibuprofen from 3 × 3 ES-HFMs loaded with photoresponsive couplers over 3 irradiation cycles (irradiation 1/2/3) after 1 h of continuous irradiation at 365 nm. (C) Ibuprofen release from 3 × 3 ES-HFMs with release triggered by 3 cycles (dose 1/2/3) of exposure to light for 1 h. The 1% and 5% loading are ES-HFMs containing 1% and 5% light-responsive couplers. Copied with permission [42]. Copyright 2016, the American Chemical Society Publications.

    Figure 3  Engineering structure of ES-HFMs. (A) Schematic diagram of the working mechanism of ES-HFMs, a patch consisting of a swellable and adhesive mussel adhesive proteins (MAP)-based shell and a non-swellable silk fibroin (SF)-based core. (B) Macroscopic images of rat skin wound healing after non-treatment, suture treatment and 60% double-layered MN (DL-MN) treatment. (C) Macroscopic images of a 5 mm diameter defect in the rat ileum after untreated, suture-treated and mussel adhesive protein/hyaluronic acid single-layered MN (MAP/HA SL-MN) treatment. Defect (black star), treated site (black dashed circle), the border between attached bowel (yellow dashed line), adhesion site indicating intestinal canal (yellow arrow), fatty attachment (black arrow) and leakage (yellow triangle). Copied with permission [32]. Copyright 2019, the Elsevier.

    Figure 4  Bioactive nanocomponents of ES-HFMs. (A) The active MN patch component, as well as the built-in Mg particles that act as a pump when in contact with body fluids, are activated, thus enhancing drug release. (B) Digital photograph showing an optical/fluorescent microscope image of a 15 × 15 MN array and an active MN tip. (C) SEM image of a single active MN tip and EDX analysis of Mg. (D) Manufacturing steps for MNs: poly(dimethylsiloxane) microforming on the primary MN, poly(dimethylsiloxane) negative MN demolding, Mg particle loading, polymer and payload inclusion, polymer drying, adhesive application and peeling. (E) Microscope time frame image is taken from a single reactive MN point, clearly showing the dissolution of the polymer in PBS pH 6.0 and the activation of the particles (at 30-second intervals). (F) Schematic diagram illustrating the experimental set-up of passive (ⅰ) and active (ⅱ) MNs penetrating pig skin, and cross-sectional images of fluorescence microscopy at different times. Copied with permission [62]. Copyright 2020, the Wiley.

    Figure 5  The typical manufacturing process of ES-HFMS via micro-modeling.

    Figure 6  Application of ES-HFMs for deep point-of-care sampling. (A) Preparation of PVA/CS hydrogel ES-HFMs. (B) Target biomarkers were extracted from simulated skin in vitro and rabbit dorsal skin in vivo. (C) Recovery of target biomarkers from ES-HFMs by heating and detection using a colorimetric assay (darker color indicated higher glucose concentration). (D) The mass of ISF extracted from rabbits by the patch increased with extraction time and reached 1.25 ± 0.37 mg after 10 min. (E) Glucose levels obtained by ES-HFMs and a commercial glucose meter were excellent consistent in terms of quality and quantity. (F) Images before and after insertion of ES-HFMs into the skin of rabbits showing no breakage despite a slightly blunted needle tip. (G) Microchannels on the skin surface of the rabbit's back disappeared rapidly within 3 min after removal of ES-HFMs. Copied with permission [67]. Copyright 2020, the Wiley.

    Figure 7  Application of ES-HFMs for smart drug deployment. (A) Schematic diagram of the application and retention strategy for rat experiments designed to assess the in vivo performance of ES-HFMs. (B) After 24 h of insertion, despite extensive swelling, ES-HFMs can still be removed intact from rat skin. (C) Digital image of a lyophilized wafer loaded with sodium ibuprofen. (D) Digital image of lyophilized wafer loaded with OVA. The plasma concentration of ibuprofen sodium (E) and OVA (F) in rats after transdermal administration using ES-HFMs. (G) Cumulative permeation profile of ibuprofen sodium in porcine skin by in vivo ES-HFMs combined with a lyophilized drug reservoir. (H) Cumulative permeation profile of OVA in porcine skin by in vivo ES-HFMs combined with a lyophilized drug reservoir. Copied with permission [68]. Copyright 2014, the Plos One.

    Figure 8  Application of ES-HFMs for wound healing. (A) Schematic representation of DL-AMNB, including biofunctional peptides fusion mussel adhesive protein for myocardial infarction treatment. (B) In vivo DL-AMNB treatment effect in a rat myocardial infarction model. Photographs of DL-AMNB treatment of the infarct zone before (ⅰ) and after (ⅱ). The yellow dashed circles represent the infarcted area. (ⅲ) Macroscopic images of the remaining DL-AMNB adhering to heart tissue 2 weeks after treatment. Copied with permission [90]. Copyright 2021, the Elsevier.

    Figure 9  Application of ES-HFMs for mechanical interlocking. (A) The mechanical interlocking of ES-HFMs after tissue penetration. (B) ES-HFMs were prepared using PDMS molds and show the chemical structure of polystyrene-block-poly-acrylic acid (PS-b-PAA), where the weight fraction of polystyrene used for the expandable tip is 25%. (C) Schematic diagram showing the internal structure and reversible water reactivity of a bilayer MN. (D) Side views of hollow MNs (without polystyrene core) with different PS-b-PAA filling ratios, including (ⅰ) 20%, (ⅱ) 40% and (ⅲ) 70%. (ⅳ) Bilayer MN with expandable tip after filling with polystyrene core. (E) Photograph of ES-HFMs at a density of 10 × 10 cm−2, showing high pattern fidelity. Copied with permission [58]. Copyright 2013, the Nature publishing group.

    Figure 10  Application of ES-HFMs for mechanical interlocking based on thorny-headed worm. (A) (ⅰ) Thorny-headed worm attaches to the intestine of fish. (ⅱ) The micro-needle-like structures in the head of the worm expand as they penetrate the intestine, providing strong adhesion. (ⅲ) After intake, the "jack-in-the-box" system would automatically deploy into the tissues in reaction to the low pH environment in the stomach. (B) Schematic diagram of a device comprising a double DOAMS modified tablet. (C) The image display device from the high-speed camera is successfully triggered. (D) Pictures of the state of the double DOAMS modified tablets before and after initiation. (E) Endoscopic images showing the movement of control and individual DOAMS-modified tablets in the pig stomach. Copied with permission [106]. Copyright 2022, the Science Advances.

    Table 1.  Typical ES-HFMs and their applications.

    下载: 导出CSV
  • 加载中
计量
  • PDF下载量:  5
  • 文章访问数:  487
  • HTML全文浏览量:  64
文章相关
  • 发布日期:  2023-06-15
  • 收稿日期:  2022-09-29
  • 接受日期:  2022-12-12
  • 修回日期:  2022-11-20
  • 网络出版日期:  2022-12-26
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

/

返回文章