引用本文:
Bing Chun XUE, Er Bao LIU. Determination of Gossypol in Trace Level by Flow Injection Analysis with Chemiluminescence Detection[J]. Chinese Chemical Letters,
2006, 17(1): 57-60.
Citation: Bing Chun XUE, Er Bao LIU. Determination of Gossypol in Trace Level by Flow Injection Analysis with Chemiluminescence Detection[J]. Chinese Chemical Letters, 2006, 17(1): 57-60.

Citation: Bing Chun XUE, Er Bao LIU. Determination of Gossypol in Trace Level by Flow Injection Analysis with Chemiluminescence Detection[J]. Chinese Chemical Letters, 2006, 17(1): 57-60.

Determination of Gossypol in Trace Level by Flow Injection Analysis with Chemiluminescence Detection
摘要:
This work reports the single-molecule detection of gossypol by flow injection analysis with chemiluminescence method. The method is based on the reaction of luminol with ferricyanid in sodium hydroxide medium sensitized by gossypol. Under the optimum conditions, the CL intensity is proportional to the concentration of gossypol over the range of 1.11×10-17-2.78×10-16 mol/L in acid solution and 8.00×10-11-7.39×10-8 mol/L in neutral solution with correlation coefficients 0.9983 and 0.9905, respectively. The detection limits is 1.60×l0-18 mol/L (S/N=3). The proposed method has been applied for the determination of the gossypol in cottonseeds and pharmaceutical formulations with satisfactory results.
English
Determination of Gossypol in Trace Level by Flow Injection Analysis with Chemiluminescence Detection
Abstract:
This work reports the single-molecule detection of gossypol by flow injection analysis with chemiluminescence method. The method is based on the reaction of luminol with ferricyanid in sodium hydroxide medium sensitized by gossypol. Under the optimum conditions, the CL intensity is proportional to the concentration of gossypol over the range of 1.11×10-17-2.78×10-16 mol/L in acid solution and 8.00×10-11-7.39×10-8 mol/L in neutral solution with correlation coefficients 0.9983 and 0.9905, respectively. The detection limits is 1.60×l0-18 mol/L (S/N=3). The proposed method has been applied for the determination of the gossypol in cottonseeds and pharmaceutical formulations with satisfactory results.
-
Key words:
- Single-molecule detection
- / chemiluminescence
- / flow injection
- / gossypol

计量
- PDF下载量: 0
- 文章访问数: 1464
- HTML全文浏览量: 6