Flow-injection Chemiluminescence Sensor for Hydrogen Peroxidewith Immobilized Reagents

Bao Xin LI Wei QIN Zhu Jun ZHANG

引用本文: Bao Xin LI,  Wei QIN,  Zhu Jun ZHANG. Flow-injection Chemiluminescence Sensor for Hydrogen Peroxidewith Immobilized Reagents[J]. Chinese Chemical Letters, 1998, 9(5): 471-472. shu
Citation:  Bao Xin LI,  Wei QIN,  Zhu Jun ZHANG. Flow-injection Chemiluminescence Sensor for Hydrogen Peroxidewith Immobilized Reagents[J]. Chinese Chemical Letters, 1998, 9(5): 471-472. shu

Flow-injection Chemiluminescence Sensor for Hydrogen Peroxidewith Immobilized Reagents

  • 基金项目:

    This study was supported by the National Natural Science Foundation of China (No. 39730160).

摘要: A novel chemiluminescence (CL) sensor for H202 combined with flow injection analysis is described in this paper. H202 is sensed by the CL reaction with luminol and Co2+ bleeding from the ion exchange column with immobilized reagents by hydrolysis. The calibration graph is linear and falls between the range of 4×l0-8 to 1×10-5 mol/L. A complete analysis could be performed in 1 min. The system was reused for over 50 hours.