Citation: Cai Shengmin, Liu Changyi, S. M. Wilhelm, Norman Hackerman. INSTRUMENTION FOR RAPID MEASUREMENT OF THE CAPACITANCE AND RESISTANCE OF ELECTROLYTIC CELL[J]. Acta Physico-Chimica Sinica, 1986, 2(01): 22-29. doi: 10.3866/PKU.WHXB19860104
电解池电阻和容抗的快速测定仪
测定了滴汞电极在1 mol L~(-1)Na_2SO_4溶液中及1 mol L~(-1)Na_2SO_4被正庚醇饱和的溶液中的微分电容~电势数值。证明仪器工作良好。另外, 从实验数据和理论分析, 均表明D.C.Grahame的被广泛引用的有关经典数据有误。本文做了修正。还测定了半导体氧化铁电极在光照下黑暗中的阻抗谱图和氧化银电极充电和停止充电后开路下的电阻~时间图, 显示了仪器在电化学研究中利用的可能性。
English
INSTRUMENTION FOR RAPID MEASUREMENT OF THE CAPACITANCE AND RESISTANCE OF ELECTROLYTIC CELL
A.C. impedance is an extremely valuable tool for the interpretation of electrochemical reaction mechanisms and
cell characteristics.
We describe here instrumentation capable of measuring the capacitance and resistance of electrochemical cells under
conditions of varying cell voltage and/or measuring frequency.
Circuit
The principle on which the circuit is based (Figure 1) is that of phase-sensitive detection and is similar to methods
described previously [1] with two exceptions. In most impedance measurement circuits, the cell voltage is modulated and the
corresponding response of the cell current monitored. This sometimes presents aproblem in separating the cell response from
that of the control circuit and generally requires larger perturbation signals. In this case, a small (<10 μA) current
signal is applied with the control circuit isolated through L. Second, in this system, the measuuring frequency can be varied
using a ramp generator coupled to a voltage-controlled frequency oscillator.
The lock-in amplifier that was used (PAR Model 5204) contains two phasesensitive detectors that are driven by ortho nal
reference signals. Both cell resistance (in phase) and cell reactance (-1/ωC, quadrature) can be monitored simultaneously.
Any phase shift caused by the measurement circuit can be offset by substituting a standard resistor for the electrochemical
cell and adjusting the phase control of the lock-in amplifier. In all cases tested, this adjustment was less than ±2°.
Applications
The response of the measuring system to a dummy cell consisting of a 50 Ω resistor in series with a 50 Ω resistor and 1 μF
capacitor in parallel gave a spectrum that agreed with the theoretical prediction within 2%. In Figure 2, the differential
capacitance of a dropping mercury electrode in 1 molL~(-1) Na_2SO_4 solution obtained using the system under discussion is
compared with Grahame′s original data [2,3]. Curve b is for a solution saturated with N-heptyl alcohol. The agreement is
quite od with the exception of the adsorption peaks. The peak heights should be sensitive to the perturbation signal
magnitude in that higher amplitude perturbations tend to “average out” sharp peaks. The small current signal that was used
in the present case seems to improve the resolution of the capacitance peaks.
There an error in the original report by Grahame(2). The abscissas in Figures 20 and 21 should read, “E relative to the
normal mercury/mercurous sulfate electrode” instead of “E relative to the normal calomel electrode”. The reasons are as
follows:
1) Our numerous experiments using SCE reference electrode and normal mercury/mercurous sulfate reference electrode show
Grahame′s abscissa (in fig.21) is incorrect.
2) The related fig.20 of Grahame′s paper was quoted from uy (Ann. Chim. Phys. 8, 291). However, we found out in this
original paper uy used a large mercury pool asreference electrode. Since it is a normal sulfate solution, the
electrodepotential should related to mercury/mercurous sulfate electrode but not to NCE electrode.
3) From table 1 on page 451 in Grahame′s paper, the E~(MAX) (potential of the electrocapillary maximum) in Na_2SO_4 solution
is -0.48 V versus NCE. It agrees well with Frumkin′s data -0.47 V versus NCE (A.H.фрумкин et.al. КИНЕТИКА Э
ЛЕКТРОДНЫХ ПРОЦЕССОВ. 1952, p.31, table 1). If in fig.20 the maximum is -0.85 V “relative to normal
mercury/mercurous sulfate electrode”, since N-mercury/mercurous electrode is 0.335 V more positive that NCE, now the maximum
appears at -0.515 V releative to NCE electrode. It agrees well with other data.
Our intention of pointing out this minor error in Grahame′s paper is merely because this paper is generally accepted as the
classical paper with most authority in its own field and Figure 21 is widely quoted in numerous papers and text-books (e. g.
A. J. Bard, L. R. Faulkner, “Electrochemical Methods, fundamentals and Applications.” 1980. p.550, 551).
The performance of silver oxide electrodes is knowng to be sensitive to the magnitude of the charging current (4). Upon
chargin at 17 mA/cm~2 (Curve a, Figure 3), potential plateaux corresponding to
2Ag+2OH~-→Ag_2O+H_2O+2e~-
and
Ag_2+4OH~-→Ag_2O_3+2H_2O+4e~-
are observed prior to oxygen evolution (highest plateau). The potential decay curve displays three plateaux corresponding to
the reverse reactions
Ag_2O_3→A →Ag_2O→Ag
In concentrated electrolyte, the cell resistance reflects primarily the resistance of the surface layer. Verification of this
fact was achieved by measurement of the resistance during potential decay. The initial decay of Ag_2O_3 to A proceeds at
low resistance in accordance with the relatively high conductance of these compounds Ag_2O_3, 2×10~2 Ω~(-1)cm~(-1); A ,
10~(-1) Ω~(-1)cm~(-1)). The transformation of the surface layer from A to Ag_2O proceeded concurrently with an increase in
resistance (Ag_2O3, 10~(-8) Ω~(-1)cm~(-1)). The resistance diminished upon discharge of Ag_2O to Ag. Rapid measurements of
this type facilitate successfull evaluation of Ag battery characteristics.
Complex plane spectra of iron oxide electrodes (s) were generated by continuous variation of measurement frequency during
experiments with both illuminated and non-illuminated electlodes. Analysis of the curves (Figure 4) provided an equivalent
circuit that explained the observed results. It was found that the internal resistance of the semiconducting oxide (R_(SC))
diminished under illumination simultaneousry with an increase in space charge capacitance. The capacitance and resistance
data obtained show promise toward routine evaluation of semiconducting materials and may facilitate optimization of
photoelectrochemical devices.
(This paper has been presented on 161st meeting of the Electrochemical Society, 1982, Montreal, Canada)
扫一扫看文章
计量
- PDF下载量: 1886
- 文章访问数: 2801
- HTML全文浏览量: 31

下载: